GDYD-63A Versatile AC/DC Hipot Test Set for Comprehensive Electrical Testing

GDYD-63A Versatile AC/DC Hipot Test Set for Comprehensive Electrical Testing

Brief Description:

GDYD-63A AC/DC Hipot Test Set is composed of automatic control unit and AC oil testing transformer (HV unit) which output AC voltage smoothly. Meanwhile, it has many monitoring and protection functions. When the short circuit rod of testing transformer is taken out, the output is DC voltage.


Product Detail

Product Tags

Features

• 10inch color LCD screen, with PLC automation control.

• HV Voltage, HV current, LV voltage, LV current measured at the same time, with high precision sensor and high-performance collection chip.

• The human-machine interface adopts full touch and partial key operation mode, intelligent working process, optional automatic mode, and manual mode.

• Display HV voltage, HV current, LV voltage, LV current, time, and withstanding result.

• Over-voltage, over-current and flashover protection. Output voltage, HV protection current, LV protection current (hardware has been set), mA meter protection current and timing time can be set.

• Support stage boosting (5 stages), and the voltage value of each stage can be arbitrarily set to facilitate the detection of test object problems.

• Zero-starting function. Tests can be proceeded after the voltage is back to zero.

• Store up to 8000pcs data in 3months.

• When setting voltage arrives, it will be timing automatically. After timing, voltage will be back to zero.

• When the LV current is oversetting value, the voltage will be cut off automatically and it will be back to zero. When the HV current is overset, the test object will be cut off and test can be proceeded continuously.

• With anti-interference ability, suitable for electromagnetic environment.

• Automatic error diagnosis, easy to find and solve problems, simple and convenient maintenance of equipment.

Specifications

 A. Control unit specifications

• Input voltage: AC220V±10%, 50Hz±1%

 Output capacity: 3kVA

 Low voltage output: AC 0~250V

 Output current: 15A (Max)

• Low voltage measurement range: 0-300V resolution: 0.1V

• Low-voltage current measurement range: 0.2-15A, resolution: 0.1A

• High voltage measurement range: AC 0-60kV, resolution: 0.01KV

• High voltage current measurement range: AC 1.0-50mA, resolution: 0.01mA

 Milliamp meter current measurement range: 0-20mA, resolution: 0.01mA

• Timing range: 0-999s

• Voltage accuracy ≤±3.0%FS

 Current accuracy ≤±3.0%FS

 B.Testing transformer (HV unit) specifications

 Output capacity: 3kVA

 Low voltage (input) voltage: AC 0~200V

 Low voltage (input) current: AC 0-15A

 High voltage (output) voltage: AC 0-60kV

• High voltage (output) current: AC 0-50mA

 Measurement ratio: 600:1 Error: ±1%

 Impedance voltage: ≤12%.

 No-load current: ≤10%.

• Running time: 5min open at rated capacity, 5min off continuous cycle duty system, long-term operation at 1/3 rated voltage and 1/3 rated current. 

C. Working conditions.

 Power supply voltage: 220V±10%, 50HZ±1%, the waveform distortion rate is not more than 5%.

• Ambient temperature: -5℃-40℃

• Ambient humidity: ≤80%RH, no condensation, no flickering of the instrument

• Altitude: ≤1000m

• The working environment should be free from obvious dust, smoke, corrosive, flammable gas, water vapor and salt spray, and no serious vibration.


  • Previous:
  • Next:

  • Send your message to us:

    Write your message here and send it to us

    Send your message to us:

    Write your message here and send it to us